HSG-C - Heat-Sealing Mashine

Test the seam strength of the sealed samples with our Universal Tensile Tester VNG-E.

An optional Hot Tack Device can be obtained as an accessory.

Also, an optional load cell enables you to adjust the force display any time. You may adjust the Temperature, Force and Time parameters with our newly developed HSG-P testing device.

Download Datasheet

Specifications  
Test method
Simulation of a head sealing equipment to determine the best sealing parameters
Compressed-air supply
max. 10 bar
Seal jaws
150 x 100 mm² smooth(standard). Other specifications upon request
Seal temperature
From ambient temperature up to 300°C max. (top and bottom sealing jaws can be separately adjusted)
Tolerance
< 1.5% FS
Sealing force
40N up to 990N for the entire sealing area
Tolerance < 1.5% FS
Dwell time
0.1 - 50.0 s
Tolerance0.3% FS
PC interface
serial standard interface RS232, 9.600 Baud
Interface for force adjustment
Internal protocol
Dimensions
68 x 35 x 44 cm
Weight
24kg
Storage temperature
0°C - 50°C
Working temperature
15°C - 35°C
Relative humidity
max. 80%, non-condensing
Electrical connection HSG-C
230 V / 50 - 60Hz, power consumption ca. 500W, approx.
Standard
Following ASTM F 2029 - 00

Accessories

Required Accessories
Optional Accessories
Laboratory Requirements
  • Mains Connection
  • Compressed-Air connector ( 6 - 10 bar )
Expendables and recommended spares
  • PT 100
  • Sealing jaws
  • Fixing Springs for the sealing unit
  • Heating cartridge
  • fuses